L'ÉTS vous donne rendez-vous à sa journée portes ouvertes qui aura lieu sur son campus à l'automne et à l'hiver : Samedi 18 novembre 2023 Samedi 17 février 2024 Le dépôt de votre demande d'admission à un programme de baccalauréat ou au cheminement universitaire en technologie sera gratuit si vous étudiez ou détenez un diplôme collégial d'un établissement québécois.

Healthcare Technology Mechanical Engineering Research and Innovation Health Technologies LIO – Imaging and orthopedics research laboratory

Toward the Design of a Prosthetic Ligament for Cattle

cow

Image purchased on Istock and protected by copyright.

SUMMARY

Cranial cruciate ligament injury is infrequent but disabling in adult cattle. It requires a surgical intervention using ligament substitutes which, unfortunately, are still inadequate because of their lack of workability and a lack of documentation on the actual mechanical properties of the bovine cranial cruciate ligament. Therefore, this research study aims to assess the quasi-static tensile mechanical properties of the healthy ligament in order to develop design specifications for a new substitute. Then, a NiTiNOL filament braid prototype of a prosthetic ligament was designed, manufactured, and tested. Results support the viability of this project and also pave the way for new surgical applications in large and small animals.

Rupture of the Anterior Cruciate Ligament

Anterior cruciate ligament rupture is a common, documented, and easily treated injury in humans [1]. Mainly associated with sports, its occurrence is often overlooked in animals. In cattle, the cranial cruciate ligament (CrCL)—equivalent to the human anterior cruciate ligament (Figure 1)—accounts for 21% of direct cases of instability in the stifle joint—equivalent to the human knee [2].

Anterior cruciate ligament

Figure 1 Torn anterior cruciate ligament in the human knee

The first signs of a torn CrCL appear in an animal as a lameness, followed by an inability to move, then to get up. CrCL rupture leads to rapid degradation of the animal’s general health and compromises its ability to feed itself [3]. With such suffering, the animal can no longer be used for milk production [4, 5] or for reproduction, causing economic loss for producers [6].

Inadequate Synthetic Substitutes

Given farmers’ natural reluctance to slaughter one of their animals, coupled with the fact that this would be an economically unprofitable situation, veterinary surgeons are called upon to stabilize the stifle joint. The surgical procedure, called the “Hamilton Technique”, consists of piercing the femoral and tibial extremities of the animal to pass a ligament substitute through [7]. Currently preferred are synthetic substitutes such as a nylon monofilament or, in other words, a heavy gauge fishing line.

Unfortunately, these materials are difficult to implant and their mechanical properties do not match the few documented mechanical properties of bovine CrCL [8, 9], leading to failure in half of the surgeries. Moreover, the documented properties are inconsistent and insufficient in establishing viable design specifications for an alternative prosthesis.

The objective of this research project carried out in collaboration with the Faculté de médecine vétérinaire de Montréal, was to assess the quasi-static tensile properties of a healthy CrCL, then to establish the preliminary design of a substitute replicating these properties.

Assessment and Mathematical Modeling

The quasi-static tensile properties assessment of a healthy CrCL was performed on 14 stifles from 7 adult female cows. These animals had all been previously euthanized at the Faculté de médecine vétérinaire for serious health disorders that had no impact on the structure and mechanical properties of their ligaments. Their CrCLs were first mechanically preconditioned and then subjected to a controlled tension until failure. Digital image correlation technology was used to adequately characterize the ligament, its local behavior, and its failure mechanisms. Analysis of the results made it possible to establish a master curve for the mechanical behavior of a healthy CrCL, and an experimental corridor, creating a guide for the design of a new ligament substitute (Figure 2). Since the mechanical properties of a ligament did not appear to be influenced by bovine age, weight, or the hindlimb from which the ligament was extracted, one substitute model can be used in all the circumstances of the present study [10, 11].

Strain rate tests on CrCL substitutes

Figure 2 Load-deformation diagrams of the healthy CrCL, superelastic braid, and nylon monofilament (adapted from [11])

 

Mathematical modeling showed that a superelastic tubular braid of NiTiNOL (equiatomic titanium and nickel alloy) 0.39 mm-diameter filaments, with an inner diameter of 4 mm and braided at an angle of 43° makes it possible to replicate the mechanical properties of a healthy CrCL (Figure 3). A prototype was manufactured and tested in quasi-static tension. The load-deformation diagram of this test closely follows the shape of the CrCL master curve, remaining within its experimental corridor, thus demonstrating the feasibility of such a project (Figure 2). In addition, thanks to the superelastic braid low bending stiffness, when compared to a nylon monofilament, a potential revision of the surgical technique is possible [11].

 

NiTiNOL braid

Figure 3 Manufacturing process of the superelastic braid

The designed and manufactured substitute paves the way to new viable solutions for the treatment of CrCL rupture in adult cattle and could soon be implanted in animals. Ultimately, this substitute could also be adapted to meet similar needs in other animals, like cats and dogs, and could become a common tool for these applications in the veterinary medical field. Since synthetic substitutes for human cruciate ligaments are still showing some shortcomings [12], this work can also offer leads for further exploration. Finally, this project is fuelling an interest in superelastic braids for medical applications presented by École de technologie supérieure in previous work on suture threads [13, 14] and, given the versatility shown in these various research studies, other applications could emerge.

Additional information

For more information on this research, please refer to the following article and thesis:

Diotalevi, Lucien; Petit, Yvan; Brailovski, Vladimir; Nichols, Sylvain; Marchionatti, Emma; Wagnac, Éric. 2018. “Quasi-static tensile properties of the cranial cruciate ligament (CRCL) in adult cattle: towards the design of a prosthetic CrCL”. Journal of the Mechanical Behavior of Biomedical Materials. Volume 79, Pages 239–245.

Diotalevi, Lucien. 2017. “Conception préliminaire d’un câble superélastique pour la stabilisation du grasset des bovins adultes suite à une rupture du ligament croisé crânial”. Master’s thesis in Engineering, Montréal, École de technologie supérieure, 153 p.

About the authors
Lucien Diotalevi graduated from École de technologie supérieure (M.Sc.A in Engineering) and from École Nationale Supérieure des Arts et Métiers (engineering degree). His interests focus on biomechanics and biomaterials.
Yvan Petit is a professor in the Department of Mechanical Engineering at ÉTS. His research interests include computer-aided design, biomechanics, medical and protective devices, and additive manufacturing techniques.
Vladimir Brailovski is a professor in the Department of Mechanical Engineering at ÉTS. He specializes in the design and manufacture of shape memory alloy devices and process engineering for additive manufacturing.
Éric Wagnac is a professor in the Department of Mechanical Engineering at ÉTS. His research interests include biomechanics, computer design, finite element simulation, protective devices and surgical tools.